Scope
Where this method applies
Use for compatible liquid samples and for root-zone extracts only when the extraction method is defined and kept consistent.
Root-zone measurement · THC SOP
Generate consistent electrical-conductivity readings while preserving the sample method, temperature context, and instrument verification needed for comparison over time.
Scope
Use for compatible liquid samples and for root-zone extracts only when the extraction method is defined and kept consistent.
Frequency
Verify or calibrate at the interval recommended by the meter manufacturer and after storage, cleaning, probe replacement, or suspicious readings.
Preparation
Procedure
Step 1
Inspect and rinse the conductivity cell or probe according to its instructions.
Record: Note visible residue, cleaning, or probe condition.
Step 2
Measure the specified conductivity standard and complete calibration if the instrument or SOP requires it.
Record: Record standard value, observed value, and pass/fail status.
Step 3
Mix or collect the liquid sample consistently and avoid contaminating it with standard solution or residue from another sample.
Record: Record sample identity and collection method.
Step 4
Place the probe correctly in the sample, allow the reading and temperature compensation to stabilize, and record the displayed EC unit.
Record: Record EC, units, temperature if available, and time.
Step 5
Rinse the probe and store it according to its manufacturer instructions.
Record: Record exceptions or maintenance.
Quality control
Reference standard is within the instrument's stated acceptance range.
Units are recorded with every value.
Sampling/extraction method is unchanged when comparing a time series.
Evidence record
Interpretation boundary
EC reflects total ionic conductivity, not the concentration of each nutrient ion.
Different meters, units, extraction methods, and temperature handling can make numbers non-comparable unless documented.
Connected learning