Scope
Where this method applies
Use for fixed or portable air temperature and RH sensors used in plant environments.
Environment measurement · THC SOP
Document sensor placement, agreement, and operating context so temperature and relative-humidity records can be interpreted as measurements rather than assumed room truth.
Scope
Use for fixed or portable air temperature and RH sensors used in plant environments.
Frequency
Check after installation, relocation, suspected drift, exposure to condensation or contamination, and at a routine interval appropriate to the sensor and monitoring importance.
Preparation
Procedure
Step 1
Record sensor height, location relative to canopy or air sources, shielding, and any nearby heat or moisture source.
Record: Record location and placement description.
Step 2
Check housing, vents, cables, batteries, and visible contamination or condensation.
Record: Record condition and maintenance.
Step 3
Colocate or otherwise compare against the defined reference method under a stable condition.
Record: Record sensor and reference temperature/RH values and time.
Step 4
Compare observed differences to the tolerance appropriate to the equipment and use case; do not silently offset data without a documented procedure.
Record: Record difference and pass/action decision.
Step 5
Restore the sensor to its documented measurement location and confirm data logging resumes.
Record: Record completion and any configuration change.
Quality control
Sensor location is documented and repeatable.
Reference comparison has date/time and both values.
Any correction, replacement, or maintenance action is traceable.
Evidence record
Interpretation boundary
Relative humidity is temperature-dependent; a sensor in a different microclimate can disagree without either sensor being defective.